mini ETD - Certificates
Go
Leakage Control
The mini ETD provides accurate leakage test results via the use of a microprocessor. In addition, the mini ETD's rinse connectors are specially designed to eliminate dead zones where debris could remain.

A leakage test controlled by a microprocessor checks the endoscope in just over a minute

The sensitivity of the tester allows the detection of pressure losses as low as 5mbar - thus every minimal reduction in pressure (indicating a leakage) can be detected
After the leakage test, a pressure of 140mbar is constantly kept to and monitored during the entire reprocessing cycle preventing water from entering the endoscope
Rinse connecters in the mini ETD ensure all channel in the endoscope are connected allowing cleaning and disinfection solutions to reach all areas inside flexible endoscopes